← Back to recalls

TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087

Class II
Recalling firm
Baxter Healthcare Corporation
Reason for recall
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
Affected products (lot / model / serial)
Serial numbers: 107971925; UDI-DI number: 00887761974067
Units affected
1 unit
Distribution pattern
Worldwide - US Nationwide distribution.
Firm location
Deerfield, IL, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0354-2027
Recall initiated
September 29, 2026
FDA classification date
November 3, 2026
FDA report date
November 11, 2026
FDA event ID
100117
Last updated on HTM Vibe
November 17, 2026
View on openFDA