← Back to recalls

MR 5300 Model Number (REF): 1. 782110, 2. 782152;

Class II
Recalling firm
Philips North America Llc
Reason for recall
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Affected products (lot / model / serial)
1. Model Number (REF): 782110; UDI-DI: 884838099364; Serial Numbers: 63383, 63384, 63385, 63386, 63422, 63435, 63643, 63346, 63021, 63054, 63091, 63092, 63113, 63313, 63314, 63315, 63322, 63521, 63524, 63565, 63572, 63593, 63604, 63639, 63694, 63697, 63705, 63706, 63707, 63716, 63721, 63727, 63736, 63737, 63739, 63743, 63750, 63753, 63754, 63755, 63769, 63773, 63617, 63560, 63038, 63484, 63023, 63064, 63066, 63075, 63086, 63149, 63161, 63183, 63194, 63195, 63198, 63199, 63308, 63319, 63331, 63334, 63336, 63364, 63371, 63412, 63425, 63429, 63430, 63452, 63515, 63525, 63609, 63615, 63633, 63636, 63638, 63666, 63668, 63675, 63687, 63689, 63690, 63695, 63696, 63729, 63730, 63742, 63763, 63765, 63774, 63787, 63806, 63829, 63686, 63193, 63257, 63431, 63583, 63592, 63123, 63155, 63474, 63576, 63416, 63748, 63337, 63456, 63046, 63157, 63357, 63569, 63000, 63325, 63504, 63506, 63104, 63110, 63115, 63186, 63248, 63250, 63252, 63343, 63344, 63586, 63785, 63820, 63821, 63328, 63664, 63771, 63457, 63070, 63356, 63011, 63013, 63022, 63028, 63083, 63095, 63107, 63202, 63292, 63365, 63432, 63490, 63499, 63500, 63594, 63031, 63217, 63219, 63220, 63221, 63271, 63309, 63025, 63125, 63211, 63451, 63508, 63541, 63559, 63623, 63745, 63019, 63034, 63058, 63084, 63168, 63218, 63374, 63443, 63510, 63057, 63063, 63493, 63513, 63564, 63612, 63684, 63719, 63812, 63366, 63372, 63026, 63074, 63093, 63099, 63109, 63111, 63112, 63130, 63169, 63175, 63176, 63177, 63178, 63188, 63210, 63223, 63224, 63228, 63229, 63240, 63312, 63333, 63341, 63342, 63389, 63408, 63413, 63428, 63466, 63483, 63487, 63523, 63531, 63532, 63533, 63557, 63566, 63567, 63571, 63596, 63598, 63602, 63614, 63618, 63619, 63632, 63650, 63665, 63672, 63673, 63677, 63732, 63749, 63035, 63041, 63043, 63047, 63062, 63090, 63100, 63128, 63136, 63140, 63141, 63150, 63152, 63160, 63162, 63172, 63173, 63182, 63222, 63238, 63244, 63253, 63258, 63363, 63367, 63370, 63393, 63394, 63126, 63323, 63558, 63080, 63699, 63514, 63388, 63410, 63411, 63723, 63311, 63375, 63507, 63159, 63396, 63122, 63068, 63433, 63001, 63784, 63679, 63409, 63756, 63762, 63434, 63259, 63379, 63277, 63301, 63187, 63464, 63758, 63759, 63098, 63134, 63326, 63355, 63418, 63561, 63562, 63624, 63032, 63036, 63069, 63139, 63437, 63438, 63439, 63610, 63700, 63708, 63709, 63710, 63733, 63738, 63746, 63786, 63103, 63164, 63534, 63563, 63589, 63631, 63127, 63603, 63720, 63482, 63489, 63494, 63498, 63547, 63555, 63106, 63108, 63402, 63475, 63698, 63775, 63776, 63822, 63825, 63273, 63424, 63626, 63397, 63398, 63445, 63006, 63008, 63009, 63015, 63016, 63024, 63030, 63042, 63044, 63049, 63053, 63105, 63135, 63138, 63151, 63153, 63154, 63492, 63352, 63059, 63497, 63681, 63488, 63503, 63722, 63071, 63072, 63294, 63295, 63296, 63297, 63299, 63340, 63361, 63455, 63519, 63724, 63770, 63067, 63174, 63191, 63197, 63449, 63674, 63800, 63798, 63039, 63165, 63318, 63330, 63380, 63440, 63012, 63014, 63018, 63029, 63033, 63087, 63196, 63203, 63204, 63316, 63520, 63587, 63477, 63530, 63740, 63201, 63400, 63401, 63446, 63453, 63478, 63502, 63550, 63597, 63605, 63616, 63704, 63747, 63757, 63780, 63823, 63824, 63833, 63119, 63442, 63461, 63570, 63766, 63767, 63810; 2. Model Number (REF): 782152; UDI-DI: N/A; Serial Numbers: 69304, 69349, 69098, 69101, 69250, 69318, 69321, 69338, 69238, 69239, 69024, 69026, 69034, 69039, 69046, 69081, 69083, 69105, 69151, 69235, 69023, 69037, 69041, 69066, 69112, 69152, 69233, 69302, 69299, 69193, 69194, 69241, 69287, 69320, 69070, 69071, 69106, 69110, 69127, 69161, 69360, 69366, 69044, 69045, 69048, 69116, 69286, 69300, 69351, 69055, 69073, 69075, 69086, 69191, 69207, 69243, 69252, 69258, 69260, 69261, 69262, 69264, 69268, 69270, 69288, 69290, 69295, 69296, 69297, 69000, 69003, 69008, 69010, 69011, 69014, 69015, 69017, 69030, 69147, 69312, 69325, 69326, 69327, 69328, 69329, 69330, 69331, 69332, 69333, 69334, 69289, 69340, 69115, 69200, 69201, 69263, 69154, 69342, 69348, 69170, 69135, 69139, 69197, 69271, 69277, 69133, 69185,
Units affected
611
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia;
Firm location
Cambridge, MA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-3241-2026
Recall initiated
August 21, 2026
FDA classification date
September 18, 2026
FDA report date
September 30, 2026
FDA event ID
99769
Last updated on HTM Vibe
October 6, 2026
View on openFDA