- Recalling firm
- Datascope Corp.
- Reason for recall
- The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
- Affected products (lot / model / serial)
- Model: 0998-00-0802; UDI: 10607567111964; Serial Numbers: 200019347EAI, 200019847EAI, 200019447EAI, 210018625EAI, 210017925EAI, 210017525EAI, 210018425EAI, 210020144EAI, 210020944EAI, 210022544EAI, 210020244EAI, 210022144EAI, 210020044EAI, 210020544EAI, 210021344EAI, 210023244EAI, 210018025EAI, 210019025EAI, 210021144EAI, 210021044EAI, 210022344EAI, 210019225EAI, 210019125EAI, 210018525EAI, 210022844EAI, 200017647EAI, 210019725EAI, 210022744EAI;
- Units affected
- 28 units (20 US, 8 OUS)
- Distribution pattern
- Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.
- Firm location
- Mahwah, NJ, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-1587-2026
- Recall initiated
- February 6, 2026
- FDA classification date
- March 19, 2026
- FDA report date
- March 25, 2026
- FDA event ID
- 98432
- Last updated on HTM Vibe
- July 24, 2026