Blanketrol II Model 222S; Thermal regulating system.

Class II
Recalling firm
Gentherm Medical, LLC
Reason for recall
Overheating of the 115V CPU board may cause the device to shut down.
Affected products (lot / model / serial)
UDI: 1061303186165 Serial Numbers:131-2-00580 through 262-3-20277
Units affected
182 units
Distribution pattern
Worldwide - US Nationwide distribution including in the territories of Puerto Rico, and the countries of Brazil, Canada, Chile, Colombia, Costa Rica, El Salvador, Guatemala, Mexico, Republic of Panama, Saudi Arabia, Taiwan, Thailand, Bahamas, and Trinidad and Tobago.
Firm location
Blue Ash, OH, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
E-Mail
Product type
Devices
Status
Ongoing
Recall number
Z-2968-2026
Recall initiated
July 13, 2026
FDA classification date
August 14, 2026
FDA report date
August 26, 2026
FDA event ID
99478
Last updated on HTM Vibe
September 8, 2026
View on openFDA