Medline Kits containing Tego Connectors: Medline IR Pack, SKU DYNJ67205C Medline NO CATHETER DIALYSIS INSERTION, SKU CVI5200
Class II
- Recalling firm
- Medline Industries, LP
- Reason for recall
- Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
- Distribution pattern
- US Nationwide distribution.
- State
- IL
- Status
- Ongoing
- Recall number
- Z-1463-2026
- Recall initiated
- January 8, 2026