← Back to recalls

Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112

Class II
Recalling firm
Cook Incorporated
Reason for recall
The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
Affected products (lot / model / serial)
UDI-DI: 00827002161122; Lot Numbers: 16613606, 16951548, 16951548X
Units affected
97 units
Distribution pattern
US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.
Firm location
Bloomington, IN, United States
Recall type
Voluntary: Firm initiated
Product type
Devices
Status
Ongoing
Recall number
Z-0163-2027
Recall initiated
September 16, 2026
FDA classification date
October 15, 2026
FDA report date
October 21, 2026
FDA event ID
99875
Last updated on HTM Vibe
October 27, 2026
View on openFDA