Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. Description/Item: TRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX PRO Starter Kit/RE4H01P-43, PROCURE TRUE METRIX PRO Meter Only/RE4212P-00, Henry Schein TRUE METRIX PRO Meter Only/RE4099P-40, Henry Schein TRUE METRIX PRO Kit/RE4099P-43, McKESSON TRUE METRIX PRO Meter Only/RE4051P-00, Moore Medical TRUE METRIX PRO Meter Only/RE4068P-01.
Class I
- Recalling firm
- Trividia Health, Inc.
- Reason for recall
- The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
- Distribution pattern
- Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.
- State
- FL
- Status
- Ongoing
- Recall number
- Z-1546-2026
- Recall initiated
- February 6, 2026
View on openFDA