- Recalling firm
- Medline Industries, LP
- Reason for recall
- Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
- Affected products (lot / model / serial)
- UDI-DI 10198459704543 Lot 26DBB599
- Units affected
- 9,272 kits
- Distribution pattern
- US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
- Firm location
- Northfield, IL, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-2608-2026
- Recall initiated
- May 28, 2026
- FDA classification date
- June 26, 2026
- FDA report date
- July 8, 2026
- FDA event ID
- 99209
- Last updated on HTM Vibe
- July 23, 2026