Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Class II
Recalling firm
Philips North America Llc
Reason for recall
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
Affected products (lot / model / serial)
UDI: (01)00884838101111
Units affected
62
Distribution pattern
U.S. and U.S. Territories.
Firm location
Cambridge, MA, United States
Recall type
FDA Mandated
Product type
Devices
Status
Ongoing
Recall number
Z-3155-2026
Recall initiated
August 18, 2026
FDA classification date
September 29, 2026
FDA report date
October 7, 2026
FDA event ID
99865
Last updated on HTM Vibe
October 13, 2026
View on openFDA