- Recalling firm
- Medline Industries, LP
- Reason for recall
- Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
- Affected products (lot / model / serial)
- DYNJ901328G UDI-DI 10198459601538 Lots 26ABD786 26CBD768 26DBJ434 DYNJ903465K UDI-DI 10198459292743 Lots 25GDB257 26ADA464 26BDA587 26DDA021 DYNJ909090G UDI-DI 10198459406027 Lots 25LBH690 25LBP697 DYNJ906102M UDI-DI 10198459406409 Lots 25JBE315 25KBI700 25KBM827 25LBB028 26ABH106 26CBQ206 DYNJ902664M UDI-DI 10198459349126 Lots 25ILA177 25KLA500 DYNJ902664N UDI-DI 10198459675287 Lot 26BLA783
- Units affected
- 9,272 kits
- Distribution pattern
- US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
- Firm location
- Northfield, IL, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-2607-2026
- Recall initiated
- May 28, 2026
- FDA classification date
- June 26, 2026
- FDA report date
- July 8, 2026
- FDA event ID
- 99209
- Last updated on HTM Vibe
- July 23, 2026