Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder form of DBM, while Grafton DBM Flex, Putty, Matrix, Crunch" and Orthoblend are produced from a fiber form of DBM Crunch" also contains demineralized bone chips/cubes, while Grafton DBM Orthoblend contains non-demineralized cancellous bone chips in addition to DBM. Grafton DBM and Grafton Plus DBM are malleable and can be molded or cut into various sizes and shapes according to the intended implant site.

Class II
Recalling firm
Medtronic Sofamor Danek USA Inc
Reason for recall
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection
Affected products (lot / model / serial)
UDI-DI: 00643169123021. Serial Numbers: A62881-034 through A62881-051, specifically: A62881-034, A62881-035, A62881-036, A62881-037, A62881-038, A62881-039, A62881-040, A62881-041, A62881-042, A62881-043, A62881-044, A62881-045, A62881-046, A62881-047, A62881-048, A62881-049, A62881-050, A62881-051
Units affected
18
Distribution pattern
US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.
Firm location
Memphis, TN, United States
Recall type
Voluntary: Firm initiated
Product type
Devices
Status
Ongoing
Recall number
Z-0137-2027
Recall initiated
August 12, 2026
FDA classification date
October 9, 2026
FDA report date
October 21, 2026
FDA event ID
99701
Last updated on HTM Vibe
October 27, 2026