Atellica CH Urine Albumin (UAlb). Material Number: 11537225

Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.
State
NY
Status
Ongoing
Recall number
Z-1484-2026
Recall initiated
January 20, 2026