← Back to recalls

RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116

Class II
Recalling firm
Steris Corporation
Reason for recall
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Affected products (lot / model / serial)
UDI: 00724995245955 Lot numbers with expiration dates: 689171 5/4/2027 689981 5/12/2027 690993 5/26/2027 691739 6/2/2027 692570 6/11/2027 694357 7/1/2027
Units affected
11806 units
Distribution pattern
Worldwide - US Nationwide distribution.
Firm location
Mentor, OH, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0155-2027
Recall initiated
September 21, 2026
FDA classification date
October 14, 2026
FDA report date
October 21, 2026
FDA event ID
99804
Last updated on HTM Vibe
October 27, 2026
View on openFDA