Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/130; Material Number: M001195210

Class II
Recalling firm
Boston Scientific Cork, Ltd.
Reason for recall
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Affected products (lot / model / serial)
Material Number: M001195210; UDI-DI: 08714729839323; Batch Numbers: 37914435, 37914436, 37956674, 37981343, 37981344, 38022632, 38022633, 38022634, 38149673, 38181211, 38224528, 38259475, 38259476, 38290713, 38290714, 38290715, 38290716, 38290717, 38399232, 38446223, 38479515, 38506196, 38530446, 38530447, 38629332, 38629333, 38647617, 38647618, 38649501, 38649508, 38687327, 38708440, 38708441, 38745184, 38769513, 38769514, 38934430, 39189795, 39189798, 40466471, 40466472, 40482683, 40482684, 40482685, 40487838, 40487839, 40487900, 40487954, 40487955, 40487956, 40487957, 40501048, 40501049, 40514122, 40514123, 40514222, 40514223, 40533342, 40533344, 40536165, 40536166;
Units affected
2547
Distribution pattern
Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.
Firm location
Cork, Ireland
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0224-2027
Recall initiated
September 16, 2026
FDA classification date
October 22, 2026
FDA report date
October 28, 2026
FDA event ID
99906
Last updated on HTM Vibe
November 3, 2026