- Recalling firm
- Medline Industries, LP
- Reason for recall
- Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
- Affected products (lot / model / serial)
- DYNJ911135 UDI-DI 10198459264382 Lots 25EDA279 25FDB612 25LDB658 DYNJ911660 UDI-DI 10198459487071 Lots 25IMF444 25JMA458 26BMG519 26DMA914 26EMA779 26FME775 26GMA849 26GMF826 26GMM101 26GMM127 DYNJ911560 UDI-DI 10198459400261 Lots 25HMG749 25JMC886 25KMA383 25KMA392 DYNJ908720A UDI-DI 10195327423155 Lots 24DBN292 24GBJ114 24IBK504 24JBC471 25KBP733 26ABG732
- Units affected
- 3,551 kits
- Distribution pattern
- US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
- Firm location
- Northfield, IL, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-0180-2027
- Recall initiated
- August 31, 2026
- FDA classification date
- October 15, 2026
- FDA report date
- October 21, 2026
- FDA event ID
- 99888
- Last updated on HTM Vibe
- October 27, 2026