← Back to recalls

Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135 LAP CHOLE DYNJ911660 LAP NISSEN DYNJ911560 UROLOGY LAPAROSCOPIC - 51638 DYNJ908720A

Class II
Recalling firm
Medline Industries, LP
Reason for recall
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Affected products (lot / model / serial)
DYNJ911135 UDI-DI 10198459264382 Lots 25EDA279 25FDB612 25LDB658 DYNJ911660 UDI-DI 10198459487071 Lots 25IMF444 25JMA458 26BMG519 26DMA914 26EMA779 26FME775 26GMA849 26GMF826 26GMM101 26GMM127 DYNJ911560 UDI-DI 10198459400261 Lots 25HMG749 25JMC886 25KMA383 25KMA392 DYNJ908720A UDI-DI 10195327423155 Lots 24DBN292 24GBJ114 24IBK504 24JBC471 25KBP733 26ABG732
Units affected
3,551 kits
Distribution pattern
US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
Firm location
Northfield, IL, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0180-2027
Recall initiated
August 31, 2026
FDA classification date
October 15, 2026
FDA report date
October 21, 2026
FDA event ID
99888
Last updated on HTM Vibe
October 27, 2026
View on openFDA