B. Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 2.6MM PUMP SEGMENT. Model Number: B3-3632M3705.
Class I- Recalling firm
- B Braun Medical Inc
- Reason for recall
- Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
- Distribution pattern
- Worldwide distribution - US Nationwide and the country of Canada.
- State
- PA
- Status
- Ongoing
- Recall number
- Z-1802-2026
- Recall initiated
- March 19, 2026