← Back to recalls

Medline Neuro Sponges, various dimensions, labeled as: 1. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.25" X 0.25", XR, 10/PK, Medline SKU NEUROSPNG02; 2. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 0.5", XR, 10/PK, Medline SKU NEUROSPNG05; 3. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 1", XR, LF, 10/PK, Medline SKU NEUROSPNG06; 4. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 1.5", XR, LF, 10/PK, Medline SKU NEUROSPNG07; 5. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 3", XR, LF, 10/PK, Medline SKU NEUROSPNG09; 6. Neuro Sponges, Eponges neuro, Neuroesponjas, size .75" X .75", XR, LF, 10/PK, Medline SKU NEUROSPNG11; 7. Neuro Sponges, Eponges neuro, Neuroesponjas, size 1" X 1", XR, LF, 10/PK, Medline SKU NEUROSPNG13; 8. Neuro Sponges, Eponges neuro, Neuroesponjas, size 1" X 3", XR, LF, 10/PK, Medline SKU NEUROSPNG14.

Class II
Recalling firm
Medline Industries, LP
Reason for recall
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Affected products (lot / model / serial)
1. NEUROSPNG02: UDI/DI each 10197344111572, UDI/DI case 20197344111579, ALL LOTS; 2. NEUROSPNG05: UDI/DI each 20080196286659, UDI/DI case 30080196286656, ALL LOTS; 3. NEUROSPNG06: UDI/DI each 20080196286505, UDI/DI case 30080196286502, ALL LOTS; 4. NEUROSPNG07: UDI/DI each 10080196286485, UDI/DI case 40080196286486, ALL LOTS; 5. NEUROSPNG09: UDI/DI each 20080196286536, UDI/DI case 30080196286533, ALL LOTS; 6. NEUROSPNG11: UDI/DI each 20080196286550, UDI/DI case 30080196286557, ALL LOTS; 7. NEUROSPNG13: UDI/DI each 10080196286577, UDI/DI case 40080196286578, ALL LOTS; 8. NEUROSPNG14: UDI/DI each 10080196286584, UDI/DI case 40080196286585, ALL LOTS.
Units affected
4225160 units
Distribution pattern
Worldwide distribution.
Firm location
Northfield, IL, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Product type
Devices
Status
Ongoing
Recall number
Z-2008-2026
Recall initiated
March 19, 2026
FDA classification date
April 30, 2026
FDA report date
May 6, 2026
FDA event ID
98664
Last updated on HTM Vibe
July 24, 2026
View on openFDA