- Recalling firm
- HF Acquisition Co LLC
- Reason for recall
- Due to incorrect incubation process being performed on product.
- Affected products (lot / model / serial)
- Lot Code: Please see attached spreadsheet listing products that were not incubated per the following requirements : a. Spore strips used in steam sterilizers are incubated for 24 hours b. Spore strips used in chemiclave sterilizers are incubated for 72 hours c. Spore strips used in dry heat or ethylene oxide sterilizers are incubated for 7 days As noted in the previous recall scope question, HealthFirst's Field Correction scope will be expanded to include notifications to all customers in possession of 1023780 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS,1023770 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS or other products serviced by Mesa Labs.
- Units affected
- 0
- Distribution pattern
- U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, and WI.
- Firm location
- Mukilteo, WA, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-2332-2026
- Recall initiated
- April 13, 2026
- FDA classification date
- June 5, 2026
- FDA report date
- June 17, 2026
- FDA event ID
- 98695