Swan-Ganz Pacing Catheter, Models: D200F7;

Class II
Recalling firm
Becton, Dickinson and Company
Reason for recall
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
Affected products (lot / model / serial)
Model-UDI-DI/Lot(Expiration): D200F7-00690103147315/66725358(7/2/2027)
Units affected
53
Distribution pattern
Worldwide - US Nationwide distribution including in the states of TX, SC, CT, GA, VA, MO, NM, UT, NJ, CA, FL, PA, IN, TN, WA, KY, MD, WV, MI, MA, NC, MN, NY, AZ, AL, LA, CO, OK, HI, MS, IL, OH, NE, DC, KS, NV, WI, SD, IA, AR, ID, OR, WY, MT, RI, VT, DE, AK, NH, ND, ME, PR and the countries of Japan, Brazil, Colombia, Australia, New Zealand, Poland, Czech Republic, South Africa, Mauritius, Germany, Israel, Iran, Slovakia, Switzerland, Slovenia, Bulgaria, Cyprus, Saudi Arabia, United Arab Emirates (UAE), Kuwait, Malta, Bahrain, Qatar, Hungary, Morocco, Romania, Lebanon, Tunisia, Oman, Turkey, Austria, Belgium, Luxembourg, France, Netherlands, Italy, Spain, Canary Islands, Portugal, United Kingdom, Northern Ireland, Ireland, Denmark, Sweden, Iceland, Norway, Estonia, Finland, Greece, Chile, Puerto Rico, Canada.
Firm location
Irvine, CA, United States
Recall type
Voluntary: Firm initiated
Product type
Devices
Status
Ongoing
Recall number
Z-2204-2026
Recall initiated
April 8, 2026
FDA classification date
May 15, 2026
FDA report date
May 27, 2026
FDA event ID
98603
Last updated on HTM Vibe
July 24, 2026