← Back to recalls

Radius VSM ECG pre-connected Set, REF: 4842, 4695

Class II
Recalling firm
Masimo Corporation
Reason for recall
Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
Affected products (lot / model / serial)
REF/UDI-DI/Lot: 4842/10843997011003/25AJH, 25KMA, 25MSC 4695/10843997016015/24N17, 25BBJ, 25DMH, 25FBR, 24KLV, 24N17, 25AGZ, 25BBJ, 25D10, 25DMH, 25FBR, 25GJM, 25KBF
Units affected
772
Distribution pattern
Worldwide - US Nationwide distribution in the states of OK, CA , MD , MA , NY , UT , NC , TN , MN , TX , OH , PA , VA and the countries of Italy, Latvia, Iraq, Saudi Arabia, Germany, Spain, Turkey, France, Australia, Malta, Oman, United Arab Emirates, Martinique, Netherlands, Cyprus, Kuwait, Czech Republic, United Kingdom, Croatia, Switzerland, Belgium, Singapore, Romania, Colombia, Hong Kong, Israel, Finland, Poland, New Zealand, Tunisia, Chile, Algeria, Canada, Denmark, Palestinian Territory, Qatar, Greece, Paraguay.
Firm location
Irvine, CA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-2670-2026
Recall initiated
February 20, 2026
FDA classification date
July 2, 2026
FDA report date
July 8, 2026
FDA event ID
98481
Last updated on HTM Vibe
July 24, 2026
View on openFDA