← Back to recalls

HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic

Class I
Recalling firm
Instrumentation Laboratory
Reason for recall
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Affected products (lot / model / serial)
Part Number: 0009800127. UDI-DI: 08426951126545. Lot Number (Expiration Date): B40005 (01/16/2027), B38062 (11/01/2026).
Units affected
434 units
Distribution pattern
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore.
Firm location
Bedford, MA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Product type
Devices
Status
Ongoing
Recall number
Z-0001-2027
Recall initiated
September 3, 2026
FDA classification date
October 9, 2026
FDA report date
October 21, 2026
FDA event ID
99843
Last updated on HTM Vibe
October 27, 2026
View on openFDA