CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3

Class II
Recalling firm
GE Medical Systems Information Technologies Inc
Reason for recall
Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal printer paper. Ignition of the paper could potentially result in serious thermal injury.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Bangladesh, Cambodia, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, South Korea, Lebanon, Libya, Lithuania, Malaysia, Myanmar, Netherlands, New Zealand, Norway, Philippines, Poland, Qatar, Romania, Saudi Arabia, Senegal, Singapore, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom.
State
WI
Status
Ongoing
Recall number
Z-2279-2026
Recall initiated
April 24, 2026