ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)

Class II
Recalling firm
Beckman Coulter, Inc.
Reason for recall
Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
Affected products (lot / model / serial)
Serial Numbers: 24U15260 to 26U15764; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
Units affected
92 units
Distribution pattern
US Nationwide distribution.
Firm location
Indianapolis, IN, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0201-2027
Recall initiated
September 21, 2026
FDA classification date
October 21, 2026
FDA report date
October 28, 2026
FDA event ID
99882
Last updated on HTM Vibe
November 3, 2026