← Back to recalls

Impulsator, Product Number: S00010, Phasitron 5, Product Number: P5 used with Instruction Manual

Class II
Recalling firm
Percussionaire Corporation
Reason for recall
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Affected products (lot / model / serial)
Product Number/UDI-DI: S00010/00849436000389, P5/00849436000501, Instruction Manual: P20047 Rev E
Units affected
228,277
Distribution pattern
Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.
Firm location
Sandpoint, ID, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-3323-2026
Recall initiated
August 18, 2026
FDA classification date
September 28, 2026
FDA report date
October 7, 2026
FDA event ID
99617
Last updated on HTM Vibe
October 13, 2026
View on openFDA