← Back to recalls

See full list in RES, exceeds character limit 1) GYN LAPAROSCOPY CDS, Model Number: CDS981504G; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407P; 3) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 4) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 5) NHP ROBOTIC GYNE CDS, Model Number: CDS984243G; 6) ROBOTIC-LF, Model Number: CDS984543D; 7) ROBOTIC-LF, Model Number: CDS984543F; 8) ROBOTIC-LF, Model Number: CDS984543G; 9) ROBOTIC-LF, Model Number: CDS984543I; 10) GYN CDS, Model Number: CDS984860F; 11) GYN LAP PELVI, Model Number: CDS985423G; 12) GYN KIT, Model Number: DYKM1657A; 13) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL100; 14) KIT ROBOTICS GYN, Model Number: DYKMBNDL109; 15) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116D; 16) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116F; 17) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119; 18) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119A; 19) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168; 20) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168A; 21) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184D; 22) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91;

Class II
Recalling firm
Medline Industries, LP
Reason for recall
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Distribution pattern
Worldwide - US Nationwide distribution.
State
IL
Status
Ongoing
Recall number
Z-2109-2026
Recall initiated
January 7, 2026