- Recalling firm
- Medtronic MiniMed, Inc.
- Reason for recall
- For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
- Affected products (lot / model / serial)
- Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0
- Units affected
- 3,465
- Distribution pattern
- Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden.
- Firm location
- Northridge, CA, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-2778-2026
- Recall initiated
- June 16, 2026
- FDA classification date
- July 23, 2026
- FDA report date
- July 29, 2026
- FDA event ID
- 99228
- Last updated on HTM Vibe
- August 4, 2026