Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app) Model/Catalog Number: 1015472

Class II
Recalling firm
Tandem Diabetes Care, Inc.
Reason for recall
A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active target glucose value; the deferred ( Later ) portion is delivered, but the total insulin delivered may be less than intended, and the application does not provide a distinct alert that the immediate portion was omitted, potentially resulting in unrecognized under-delivery of insulin and hyperglycemia.
Affected products (lot / model / serial)
Lot Code: UDI-DI: 00389152547244 Software versions: 3.5.4, 3.5.2.2, and 3.5.2.1
Units affected
4086 users
Distribution pattern
Worldwide - US Nationwide and the country of Czech Republic.
Firm location
San Diego, CA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
E-Mail
Product type
Devices
Status
Ongoing
Recall number
Z-0437-2027
Recall initiated
October 13, 2026
FDA classification date
November 12, 2026
FDA report date
November 18, 2026
FDA event ID
100091