SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Class II
Recalling firm
Howmedica Osteonics Corp.
Reason for recall
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Affected products (lot / model / serial)
73-2710 UDI-DI 07613327033847 Lot NPPV 73-2910 UDI-DI 07613327033854 Lot 14AGFD 73-3110 UDI-DI 07613327033861 Lots 1L45 4JM5 7EL5 AME3 MK74 NTV7 PYHM RE7J VEWJ 73-3508 UDI-DI 07613327033885 Lots HMM5 NLRE 73-3708 UDI-DI 07613327033892 Lots 3467 3D8D PMRHX1 T6YJ T9ME VWDA YAM7 73-3710 UDI-DI 07613327033908 Lots 3W25 DLYP 73-3910 UDI-DI 07613327033915 Lots 14NEYA 1XED 9Y3T D1M9 D2YP EX8T KRT3 KRT3X1 T325 V8X8 X6YE
Units affected
12,156 US
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Firm location
Mahwah, NJ, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-3206-2026
Recall initiated
July 30, 2026
FDA classification date
September 16, 2026
FDA report date
September 23, 2026
FDA event ID
99558
Last updated on HTM Vibe
September 29, 2026
View on openFDA