VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Class I
Recalling firm
Ventec Life Systems, Inc.
Reason for recall
Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk
Affected products (lot / model / serial)
UDI-DI (REF/UDI-DI): PRT-01198-000/00855573007914, PRT-01185-000/00855573007877 Package/Device Serial Number (package serial number may include a "P" at the end): 152884, 152885, 152886, 152887, 152888, 152889, 152890, 152891, 152892, 152893, 152895, 152896, 152897, 152898, 152899, 152900, 152901, 152902, 152903, 152904, 152908, 152909, 152910, 152911, 152912, 152913, 152914, 152915, 152916, 152917, 152918, 152919, 152986, 152987, 155236, 156507, 156512, 156515, 156516, 156517, 156524, 156532
Units affected
42
Distribution pattern
US: OH, GA, CA, MO
Firm location
Bothell, WA, United States
Recall type
Voluntary: Firm initiated
Product type
Devices
Status
Ongoing
Recall number
Z-1942-2026
Recall initiated
March 25, 2026
FDA classification date
May 1, 2026
FDA report date
May 13, 2026
FDA event ID
98554