← Back to recalls

Brand Name: Quantum Rehab Product Name: Quantum 4Front 2HD Model/Catalog Number: Quantum 4Front 2HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

Class II
Recalling firm
Pride Mobility Products Corp.
Reason for recall
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Affected products (lot / model / serial)
Lot Code: Quantum 4Front 2HD 0060650920020*2 K143383 D246699 41875 ITI
Units affected
1102
Distribution pattern
Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.
Firm location
Duryea, PA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-2851-2026
Recall initiated
June 25, 2026
FDA classification date
July 31, 2026
FDA report date
August 12, 2026
FDA event ID
99286
Last updated on HTM Vibe
August 18, 2026
View on openFDA