Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431K; 3) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431O; 4) KIT ROBOTICS GYN, Model Number: DYKMBNDL109A; 5) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116A; 6) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116G; 7) ROBOTIC PACK, Model Number: DYNJ04135B; 8) MINOR LITHOTOMY PACK, Model Number: DYNJ04382J; 9) NBMC D & C PACK-LF, Model Number: DYNJ0665843C; 10) D&C PACK-LF, Model Number: DYNJ22567G; 11) D&C PACK-LF, Model Number: DYNJ22567J; 12) D AND C, Model Number: DYNJ41272B; 13) D AND C, Model Number: DYNJ41272C; 14) OB PACK, Model Number: DYNJ42884; 15) D&C/GYN PACK, Model Number: DYNJ47713A; 16) PERI GYN PACK, Model Number: DYNJ55377D; 17) ANTERIOR CERVICAL-LF, Model Number: DYNJ58344; 18) D&C PACK, Model Number: DYNJ67214D; 19) D&C PACK, Model Number: DYNJ67214F; 20) D&C PACK, Model Number: DYNJ67214G; 21) D&C PACK, Model Number: DYNJ67214I; 22) LITHOTOMY PACK, Model Number: DYNJ80765C; 23) GYN PACK, Model Number: DYNJ83475; 24) D AND C PACK, Model Number: DYNJ87468; 25) D&C-ASC, Model Number: DYNJ900465C; 26) D&C-ASC, Model Number: DYNJ900465D; 27) D&C-ASC, Model Number: DYNJ900465F; 28) GYN LITHOTOMY-LF, Model Number: DYNJ902711L; 29) GYN LITHOTOMY-LF, Model Number: DYNJ902711N; 30) D&C, Model Number: DYNJ906947C; 31) PACK PERI GYN DSMC, Model Number: DYNJT6424
Class II
- Recalling firm
- Medline Industries, LP
- Reason for recall
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Distribution pattern
- Worldwide distribution - US Nationwide.
- State
- IL
- Status
- Ongoing
- Recall number
- Z-1419-2026
- Recall initiated
- January 7, 2026
View on openFDA