← Back to recalls

TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.

Class II
Recalling firm
Howmedica Osteonics Corp.
Reason for recall
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Affected products (lot / model / serial)
Catalog Number: 6543-6-519. GTIN: 07613327555127. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Units affected
577 trays
Distribution pattern
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Firm location
Mahwah, NJ, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0167-2027
Recall initiated
September 16, 2026
FDA classification date
October 15, 2026
FDA report date
October 21, 2026
FDA event ID
99883
Last updated on HTM Vibe
October 27, 2026
View on openFDA