- Recalling firm
- Daavlin Distributing Company
- Reason for recall
- Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Affected products (lot / model / serial)
- Serial Numbers: 4-01440NB10CL, 2 4-01441NB10CL.
- Units affected
- 2 units
- Distribution pattern
- US Nationwide distribution.
- Firm location
- Bryan, OH, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Telephone
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-2742-2026
- Recall initiated
- March 5, 2026
- FDA classification date
- July 15, 2026
- FDA report date
- July 22, 2026
- FDA event ID
- 99331
- Last updated on HTM Vibe
- July 28, 2026