- Recalling firm
- Beckman Coulter, Inc.
- Reason for recall
- Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
- Affected products (lot / model / serial)
- Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
- Units affected
- 336 units
- Distribution pattern
- US Nationwide distribution.
- Firm location
- Indianapolis, IN, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-0202-2027
- Recall initiated
- September 21, 2026
- FDA classification date
- October 21, 2026
- FDA report date
- October 28, 2026
- FDA event ID
- 99882