Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 37351-30-N, 37351-32-N, 37351-34-N, 37351-36-N, 37351-38-N, 37351-40-N, 37351-42-N, 37351-44-N, 37351-46-N, 37351-48-N, 37351-50-N, 37351-55-N, 37351-60-N, 37351-65-N, 37351-70-N, 37351-75-N, 37351-80-N, 37351-85-N. 2. Humeral Plate, Distal; Article Numbers: 21241-5, 21241-7, 21242-7, 21243-10, 21244-9.
Class II
- Recalling firm
- I.T.S. GmbH
- Reason for recall
- Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
- Distribution pattern
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Status
- Ongoing
- Recall number
- Z-1602-2026
- Recall initiated
- January 23, 2026
View on openFDA