- Recalling firm
- INTEGRATED MEDCRAFT LLC
- Reason for recall
- Subrecall of a kitted device from SafeGuard Medical due to internal testing identified an issue affecting the device's sterile pad, potentially causing the sterile pad to degrade over time. The degradation may cause remnants of the nonwoven pad material to dislodge on the wound or in close proximity to the wound.
- Affected products (lot / model / serial)
- Model: ATF-GSW; UDI-DI: 00850056748063; Lot Numbers: IMC-00000024, IMC-00000034, IMC-00000053, IMC-00000056, IMC-00000076, IMC-00000081, IMC-00000082, IMC-00000083, IMC-00000084, IMC-00000085, IMC-00000094, IMC-00000148, IMC-00000149, IMC-00000152, IMC-00000153, IMC-00000154, IMC-00000173, IMC-00000174, IMC-00000179, IMC-00000186, IMC-00000187;
- Units affected
- 1033
- Distribution pattern
- US Nationwide distribution in the states of GA, VA.
- Firm location
- East Hampton, CT, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-0147-2027
- Recall initiated
- September 25, 2026
- FDA classification date
- October 9, 2026
- FDA report date
- October 21, 2026
- FDA event ID
- 99948
- Last updated on HTM Vibe
- October 27, 2026