Senographe Pristina

Class II
Recalling firm
GE Medical Systems, LLC
Reason for recall
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
Affected products (lot / model / serial)
GTIN(s): 00840682118460, 00840682142052, 00840682145879, 00195278022745, 00195278276971, 00195278661999, 00195278665379, 00195278914026, 00198953109281, 10198953054311, 10195278901764.
Distribution pattern
USA and OUS.
Firm location
Waukesha, WI, United States
Recall type
FDA Mandated
Product type
Devices
Status
Ongoing
Recall number
Z-0326-2027
Recall initiated
August 28, 2026
FDA classification date
November 4, 2026