Brand Name: Olympus Everest Bipolar Cutting Forceps, 5 mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3005 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.

Class II
Recalling firm
Olympus Corporation of the Americas
Reason for recall
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Affected products (lot / model / serial)
Model: 3005; UDI: 00821925035881; Lot Number: All unexpired
Units affected
106 units
Distribution pattern
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Firm location
Center Valley, PA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-1638-2026
Recall initiated
February 25, 2026
FDA classification date
March 25, 2026
FDA report date
April 1, 2026
FDA event ID
98454