VERIFY Biological Steam Test Pack Model/Catalog Number: S3065

Class II
Recalling firm
Steris Corporation
Reason for recall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Affected products (lot / model / serial)
UDI: 10724995220300 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027
Units affected
217
Distribution pattern
Worldwide - US Nationwide distribution.
Firm location
Mentor, OH, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0074-2027
Recall initiated
August 31, 2026
FDA classification date
October 2, 2026
FDA report date
October 14, 2026
FDA event ID
99704