- Recalling firm
- Mirion Technologies (Capintec), Inc.
- Reason for recall
- Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.
- Affected products (lot / model / serial)
- Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm.
- Units affected
- 783
- Distribution pattern
- Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA, BANGLADESH, BRASIL, Brazil, CANADA, CHILE, Costa Rica, Ecuador, France, Hanoi Vietnam, Hong Kong, INDIA, ISRAEL, ITALY, Malaysia, Philippines, POLAND, Saudi Arabia, SOUTH KOREA, Spain, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIATES, UNITED KINGDOM, VIETNAM.
- Firm location
- Florham Park, NJ, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- E-Mail
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-3193-2026
- Recall initiated
- July 20, 2026
- FDA classification date
- September 16, 2026
- FDA report date
- September 23, 2026
- FDA event ID
- 99677