← Back to recalls

Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software Version: None Product Description: The restor3d Metatarsophalangeal (MTP) Implant is a sterile, single-use implant grade metal device, available in varied articulating surface coverages. The restor3d MTP implant is designed to replace the damaged articulating surface of the first metatarsal distal head. The device is intended for the treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock and integrity of the first metatarsal head. This implant can also be used in the following clinical conditions: hallux valgus or hallux limitus, hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The implant is a single use product intended to be press fit, with optional use of bone cement. The restor3d MTP Implant is comprised of a single laser powder bed fusion (LPBF) printed cobalt-chromium alloy (Co28Cr6Mo) part. The external surface of the implant that is intended to be in contact with the phalanx is mirror polished. Component: Not a component

Class II
Recalling firm
restor3d Inc.
Reason for recall
Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.
Affected products (lot / model / serial)
Lot Code: Model: 1103-161620 UDI: 00840097505244 Lot: A02979
Units affected
16
Distribution pattern
U.S. Nationwide distribution in the states of FL, IL, NC, PA, SC, and TX.
Firm location
Durham, NC, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0264-2027
Recall initiated
September 4, 2026
FDA classification date
October 22, 2026
FDA report date
October 28, 2026
FDA event ID
99940
Last updated on HTM Vibe
November 3, 2026
View on openFDA