← Back to recalls

2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.

Class II
Recalling firm
Synthes (USA) Products LLC
Reason for recall
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
Distribution pattern
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.
State
PA
Status
Ongoing
Recall number
Z-1703-2026
Recall initiated
February 25, 2026