- Recalling firm
- Fort Defiance Industries, LLC
- Reason for recall
- The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
- Affected products (lot / model / serial)
- Lot Code: All serial numbers/ GTIN: (01) 00862992000327
- Distribution pattern
- Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
- Firm location
- Loudon, TN, United States
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-2983-2026
- Recall initiated
- July 8, 2026
- FDA classification date
- August 18, 2026
- FDA report date
- August 26, 2026
- FDA event ID
- 99371
- Last updated on HTM Vibe
- September 8, 2026