← Back to recalls

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Class II
Recalling firm
Fort Defiance Industries, LLC
Reason for recall
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Affected products (lot / model / serial)
Lot Code: All serial numbers/ GTIN: (01) 00862992000327
Distribution pattern
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Firm location
Loudon, TN, United States
Recall type
Voluntary: Firm initiated
Product type
Devices
Status
Ongoing
Recall number
Z-2983-2026
Recall initiated
July 8, 2026
FDA classification date
August 18, 2026
FDA report date
August 26, 2026
FDA event ID
99371
Last updated on HTM Vibe
September 8, 2026