Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;

Class II
Recalling firm
Bayer Medical Care, Inc.
Reason for recall
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
Distribution pattern
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DE, FL, IA, IL, IN, KS, KY, LA, MA, MD, ME, MO, NC, NY, OH, OK, PA, SC, TN, TX, VA, WI and the countries of AE, BA, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, HR, IL, KW, MT, MX, NL, RO, RU, SE, ZA.
State
PA
Status
Ongoing
Recall number
Z-2296-2026
Recall initiated
April 24, 2026
View on openFDA