- Recalling firm
- Bayer Medical Care, Inc.
- Reason for recall
- Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
- Affected products (lot / model / serial)
- Catalog number: AVA 500 MPAT (Part 2 of 2) Affected material numbers: 87629007, 60729458, 86566648; Affected batch numbers: 251702, 252302, 252502, 252602, 252702, 252802, 253304, 253702, 253704, 253804, 254202, 254304, 254704, 254802;
- Units affected
- 11,300 units (1650 US, 9650 OUS)
- Distribution pattern
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DE, FL, IA, IL, IN, KS, KY, LA, MA, MD, ME, MO, NC, NY, OH, OK, PA, SC, TN, TX, VA, WI and the countries of AE, BA, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, HR, IL, KW, MT, MX, NL, RO, RU, SE, ZA.
- Firm location
- Indianola, PA, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-2296-2026
- Recall initiated
- April 24, 2026
- FDA classification date
- June 3, 2026
- FDA report date
- June 10, 2026
- FDA event ID
- 98881
- Last updated on HTM Vibe
- July 23, 2026