- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Reason for recall
- Reagent probe carryover on the Atellica CH and Atellica CI Analyzers may cause falsely elevated Hemoglobin A1c (A1c_E/A1c_H) results due to HDLC carryover (biases of 9 29% in NGSP units), and falsely depressed Direct Bilirubin 2 (DBil_2) results due to TP_2 carryover (biases of 0% to -25%), which may affect patient samples, quality control, and/or calibrator results.
- Affected products (lot / model / serial)
- 1. Siemens Material Number (SMN): 11645755 (3700 tests); UDI-DI: 00630414654010; All lots; 2. Siemens Material Number (SMN): 11537228 (7400 tests); UDI-DI: 00630414611082; All lots;
- Units affected
- 25210 units (8087 US, 17123 OUS)
- Distribution pattern
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cura¿ao, St Eus, Czech Republic, Democratic Republic Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Mexico, Morocco, Nepal, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, U.A.E., Uganda, United Kingdom , Uruguay, Uzbekistan, Vatikancity, Vietnam.
- Firm location
- Tarrytown, NY, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-0365-2027
- Recall initiated
- September 22, 2026
- FDA classification date
- November 6, 2026
- FDA report date
- November 18, 2026
- FDA event ID
- 100035
- Last updated on HTM Vibe
- November 25, 2026