Atellica CH Total Protein_2 (TP_2); Siemens Material Number (SMN): 1. 11645755 (3700 tests), 2. 11537228 (7400 tests);

Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Reagent probe carryover on the Atellica CH and Atellica CI Analyzers may cause falsely elevated Hemoglobin A1c (A1c_E/A1c_H) results due to HDLC carryover (biases of 9 29% in NGSP units), and falsely depressed Direct Bilirubin 2 (DBil_2) results due to TP_2 carryover (biases of 0% to -25%), which may affect patient samples, quality control, and/or calibrator results.
Affected products (lot / model / serial)
1. Siemens Material Number (SMN): 11645755 (3700 tests); UDI-DI: 00630414654010; All lots; 2. Siemens Material Number (SMN): 11537228 (7400 tests); UDI-DI: 00630414611082; All lots;
Units affected
25210 units (8087 US, 17123 OUS)
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cura¿ao, St Eus, Czech Republic, Democratic Republic Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Mexico, Morocco, Nepal, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, U.A.E., Uganda, United Kingdom , Uruguay, Uzbekistan, Vatikancity, Vietnam.
Firm location
Tarrytown, NY, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0365-2027
Recall initiated
September 22, 2026
FDA classification date
November 6, 2026
FDA report date
November 18, 2026
FDA event ID
100035
Last updated on HTM Vibe
November 25, 2026
View on openFDA