CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

Class II
Recalling firm
Beckman Coulter Mishima K.K.
Reason for recall
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment
Distribution pattern
US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.
State
N/A
Status
Ongoing
Recall number
Z-1553-2026
Recall initiated
January 30, 2026