← Back to recalls

Arjo Tenor mobile passive patient lift, Model Numbers KHA1000 US and KHA1010 US

Class II
Recalling firm
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Reason for recall
An internal component of the actuators in a certain production batch may not meet the required mechanical strength. This could result, in the worst case scenario, in a sudden loss of the actuator's ability to hold the load, leading to rapid and uncontrolled downward movement of the lifting arm.
Distribution pattern
US Nationwide distribution.
State
N/A
Status
Ongoing
Recall number
Z-1811-2026
Recall initiated
March 10, 2026