← Back to recalls

ARTIS icono biplane. Model Number: 11327600.

Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Potential that the system may not execute movement control commands as intended and may display the following message to the user, "Limited stand/table movement, SC."
Affected products (lot / model / serial)
Model Number: 11327600. UDI-DI: 4056869063317. Serial Numbers: 180361, 180569, 180607. Software Version VE50B.
Units affected
3 units
Distribution pattern
US Nationwide distribution in the states of CA, KY, NY, TX.
Firm location
Malvern, PA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0369-2027
Recall initiated
September 24, 2026
FDA classification date
November 6, 2026
FDA report date
November 18, 2026
FDA event ID
100014
Last updated on HTM Vibe
November 25, 2026
View on openFDA