- Recalling firm
- Abiomed, Inc.
- Reason for recall
- Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
- Affected products (lot / model / serial)
- Product Code:0042-0040-US, 0042-0040-UK, 0042-0040-EU, 0042-0040-CA; UDI: 00813502011401, 00813502012484, 00813502012217, 00813502012958; Serial No. IC8024, IC7912, IC8179, IC8216, IC8349, IC8415, IC8418, IC8529, IC8617, IC8808, IC8852, IC8906, IC9081, IC9082, IC9198, IC9227, IC9657, IC9987, IC10097, IC10098, IC10294, IC10327, IC10530, IC10531, IC10383, IC10578, IC10607, IC10549, IC10557, IC10717, IC10722, IC10769, IC10879, IC10877, IC10900, IC10926, IC11012, IC11068, IC11158, IC11168, IC11147, IC11203, IC11353, IC13018, IC13602, IC13606, IC13609, IC13610, IC13605, IC13498, IC13509, IC13506, IC13512, IC13514, IC13515, IC13517, IC13519, IC13513, IC13516, IC13534, IC13536, IC13537, IC13538, IC13535, IC13456, IC13495, IC13485, IC13575, IC13578, IC13551, IC13554, IC13548, IC13552, IC13553, IC13556, IC13558, IC13543, IC13549, IC13555, IC13559, IC13562, IC13564, IC13557, IC13563, IC13649, IC13651, IC13589, IC13591, IC13592, IC13598, IC13599, IC13600, IC13597, IC13593, IC13607, IC13608, IC13571, IC13572, IC13573, IC13596, IC13622, IC13620, IC13621, IC10025, IC9990, IC10240, IC10241, IC10592, IC10601, IC11198, IC11640, IC11642, IC11645, IC11646, IC13696, IC11123, IC11282, IC11264, IC11266, IC10146, IC10180, IC10371, IC10396, IC10397, IC10721, IC10772, IC10828, IC10835, IC11119, IC11072, IC11878, IC11958, IC11967, IC11969, IC11978, IC11986, IC11992, IC11997, IC12260, IC12261, IC12446, IC12454, IC12458, IC12459, IC12544, IC12546, IC12566, IC12586, IC12595, IC12596, IC12666, IC12670, IC12749, IC12744, IC12747, IC12768, IC10164, IC10165, IC12315, IC12318, IC9726, IC9730, IC9729, IC9830, IC10392, IC11217, IC11220, IC11215, IC11219, IC11281, IC11289, IC11952, IC11959, IC11960, IC11966, IC11970, IC11973, IC12309, IC12310, IC12307, IC12319, IC12455, IC12456, IC12467, IC12485, IC12484, IC10394, IC10400, IC10938, IC11715, IC11722, IC12220, IC12221, IC12671, IC12672, IC12561, IC12563, IC13338, IC13143, IC13144, IC13146, IC13148, IC13150, IC13151, IC13156, IC13157, IC13158, IC13142, IC8821, IC9076, IC9427, IC9766, IC10541, IC10639, IC12696, IC12700, IC12702, IC12703, IC12704, IC12753, IC12754, IC12755, IC12757, IC13281, IC10224, IC10225, IC10788, IC10789, IC10790, IC10792, IC10793, IC11723, IC11724, IC12256, IC13235, IC8836, IC8839, IC8902, IC9014, IC9022, IC10140, IC10141, IC10408, IC10425, IC10949, IC10954, IC10947, IC12320, IC12558, IC12568, IC12590, IC12597, IC12541, IC10560, IC10886, IC10892, IC10920, IC11074, IC11229, IC11998, IC12218, IC12288, IC12443, IC12444, IC12445, IC12452, IC12545, IC12552, IC12587, IC12762, IC10168, IC10125, IC11999, IC12278, IC12480, IC12543, IC12766, IC10312, IC11171, IC11172, IC12604, IC12605, IC12891, IC12895, IC12892, IC13266, IC13267, IC13503, IC13504, IC9399, IC9402, IC9626, IC10403.
- Units affected
- 293 units
- Distribution pattern
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries: Belgium, Brazil, Canada, Colombia, Croatia, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
- Firm location
- Danvers, MA, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-0210-2027
- Recall initiated
- September 18, 2026
- FDA classification date
- October 28, 2026
- FDA report date
- November 4, 2026
- FDA event ID
- 99778
- Last updated on HTM Vibe
- November 10, 2026