← Back to recalls

Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

Class II
Recalling firm
restor3d Inc.
Reason for recall
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
Affected products (lot / model / serial)
All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409
Units affected
240 units
Distribution pattern
US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.
Firm location
Durham, NC, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
E-Mail
Product type
Devices
Status
Ongoing
Recall number
Z-0348-2027
Recall initiated
October 1, 2026
FDA classification date
November 3, 2026
FDA report date
November 11, 2026
FDA event ID
100047
Last updated on HTM Vibe
November 17, 2026
View on openFDA