- Recalling firm
- restor3d Inc.
- Reason for recall
- (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
- Affected products (lot / model / serial)
- All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409
- Units affected
- 240 units
- Distribution pattern
- US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.
- Firm location
- Durham, NC, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- E-Mail
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-0348-2027
- Recall initiated
- October 1, 2026
- FDA classification date
- November 3, 2026
- FDA report date
- November 11, 2026
- FDA event ID
- 100047
- Last updated on HTM Vibe